Before setting a target number, specify who will use the product and where
The biotech and medical field in Korea's SME Strategic Technology Roadmap for 2026–2028 covers different development areas, including pharmaceuticals and biotechnology, as well as medical devices and healthcare.[1] Even the same goal of 'better performance' raises different questions for research measurements, manufacturing process management and products used in patient care. First write a sentence identifying the user, setting, inputs, outputs and the decision those outputs will support.
For example, saying that a product 'analyzes specimen images' is not enough to set an objective. The evidence needed differs between a tool for researchers comparing candidates and a tool helping medical professionals interpret patient specimens. This article helps organize a development plan; it does not determine whether a specific product is subject to regulatory approval requirements or what clinical trial requirements apply. Do not apply the medical software example below as a common testing procedure for medicines or all biotech products.
Organize supporting evidence by the question it answers
IMDRF's N41 document on the clinical evaluation of Software as a Medical Device distinguishes valid clinical association, analytical or technical validation, and clinical validation.[2] Whether an output is accurate and whether it serves an intended clinical purpose are different questions. Drawing on that distinction, a development team can organize objectives and evidence gaps as follows.
The table is IRUMJAE's proposed R&D review template. Actual evaluation methods, acceptance criteria and sample sizes must be set for the intended use and product characteristics. The table itself is not a regulator's review checklist.
| Development team's question | Objective to define | Examples of supporting evidence | What still needs checking |
|---|---|---|---|
| Why examine this output? | The relationship between a measurement or analytical result and the condition of interest | Relevant research, with the populations or subjects and conditions to which it applies | Do the literature's scope and our product's intended use match? |
| Does the input produce the required output? | Measurement or analytical performance under defined conditions | Tests against reference results, with repeat measurements and error records | Do changes in input equipment, specimens or product version affect the result? |
| Does the product achieve its purpose for the intended users and setting? | Performance and limitations to examine during product use | Evaluations designed for that purpose, with the context of use | Is the evidence needed for the users and target population available? |
Hypothetical case: do not present development scores as performance on new specimens
Imagine a team developing software to assist with the interpretation of specimen images. It obtains high scores on images repeatedly examined while building the model. It does not yet know whether those results hold for new images from another institution. Presenting scores on development data as performance in an external evaluation could lead to the wrong next development decision.
The FDA's glossary explains that test data for AI-enabled medical products should be independent of training and tuning data.[3] The team's development plan should first include checking where data can be obtained and how it may be used, then maintaining a distinction between data used for training or tuning and data reserved for final evaluation. Differences in institutions, imaging equipment and characteristics of the target population should also be considered in the evaluation plan. If the model is changed after evaluation results have been examined, record that change and the need for further evaluation.
If new data is not yet available, a plan for evaluation using independent data can be the current stage's deliverable. Separating expected performance from performance actually demonstrated is more useful for estimating the next project's costs and schedule than claiming completion too early.
Give each evidence gap a next action and an owner
For each row of the development review sheet, connect an objective ID, intended use, reference for comparison, input data and product version, test conditions, source data location, demonstrated results, unverified scope and the next decision. Instead of only writing 'improve accuracy,' specify which errors to reduce, which data will be used to assess them and how the development scope will change if results fall short.
Not every evidence gap must be closed on the same day. Decide whether to revisit the applicability of published research, restrict input equipment, or discuss evaluation methods with institutions that may be able to provide additional data. Before actual evaluations involving people, specimens or medical data, separately check the necessary rights to use them and the applicable ethics and institutional procedures. Reading international guidance does not establish compliance with Korean approval requirements.
For the role of examples and measurement conditions in patent applications, see I2B's separate article. This IRUMJAE article focuses on deciding the next R&D stage using product development objectives and validation evidence.
Next steps
To discuss organizing biotech or medical R&D evidence, send IRUMJAE a non-confidential overview of the product's intended use, current development stage, test evidence actually available and areas still unverified. Do not attach patient identifiers or non-public source data to the first inquiry.
Sources & further reading
Sources checked · 2026-11-02
[2] IMDRF/SaMD WG/N41 FINAL:2017 · Clinical Evaluation, Sections 4–5 (printed pages 8–10)
[3] FDA · Digital Health and Artificial Intelligence Glossary, Test Data